Skip to main content

LAP 324 Aerosol Spectrometer

LAP 324 Aerosol Spectrometer for analysis of particle size distributions and particle number concentrations in aerosols, Dual Wave Length Technology

The Laser Aerosol Spectrometer LAP 324 is an analytical instrument for high-resolution characterisation of both number concentration and size distribution of airborne particles.
Aerosol-Partikelgrößenspektrometer LAP 324 VorderseiteAerosol-Partikelgrößenspektrometer LAP 324 Rückseite
Aerosol particle size spectrometer LAP 324 frontsideAerosol particle size spectrometer LAP 324 backside
In accordance with ISO 21501-1, the spectrometer is based on the measuring principle of single particle light scattering using either monochromatic or polychromatic light. Therefore, the LAP 324 is equipped with two different laser diodes (red, blue) that can be operated in parallel (i.e. Dual Wave-length® technology) or individually.

The touch screen can be used for adjusting as well as operating the instrument and provides visualisation of the measured concentration course respectively the particle size distribution.

Standards

ISO 21501-1
ISO 18562-2
VDI 3867-4
IEC 63986-2-1
Interested?We will gladly advise you

+49 (351) 2166 430

Benefits

  • stand-alone or computer-assisted operation
  • high particle classification accuracy
  • wide particle number concentration range
  • operation with two or one light sources

Applications

  • high resolution analysis of aerosols
  • filter testing and filter characterisation
  • indoor aerosol analyses
  • basic aerosol research
  • toxicology testing

Within the aerosol spectrometer, the sampled aerosol flow rate is firstly divided into an effective detection flow rate and a bypass flow rate. The unmodified effective detection flow rate goes straight forward into the measurement cell. To avoid analytical artifacts, the effective detection flow rate is sheathed (aerodynamic focusing) by the afore filtered bypass flow rate.

Two long-life laser diodes of different wavelengths in the measurement cell illuminate particles passing through. The scattered light is collected by a photo detector. These signals are converted into electrical signals and processed.

Parameter title Unit Value
measurands - particle number concentration, particle size distribution
measuring range, particle size µm 0,15 ... 40
measuring range, concentration #/cm³ max. 20'000
resolution, measurement time s ≥ 1 (variable adjustable)
resolution, particle size - max 128 size channels (adjustable)
test flow rate L/min 3,0 (0,1 trough measuring cell)
power supply - 110 - 230 VAC (50 - 60 Hz); 12 VDC (4,2 A)
light source red Laser diode (660 nm, 30mW); blue Laser diode (450 nm, 60 mW)
communication interface - Ethernet, USB-C, RS232
dimensions (w × h × d) mm 220 x 380 x 200
weight kg 9,4
  • carry case
  • sample switching Unit SYS 520
  • Software PASWin (included)
  • Dilution System DIL

Similar products

LAP 323

LAP 323 Aerosol Spectrometer for analysis of particle size distributions and particle number concentrations in aerosols, Dual Wave Length Technology

Aerosol particle size spectrometer LAP 323 frontside
detail view